We design, supply, install, test and commission the isolated power systems of hospitals, and maintain the systems we installed. An isolated power system, the medical IT system of IEC 60364-7-710, supplies the medical electrical equipment of the group 2 locations: the operating theatres, the intensive care and coronary care units, the catheterisation laboratories and the anaesthetic and recovery rooms, where a sudden loss of supply would put a patient's life at risk.
Its isolating transformer separates the circuits of the room from earth, so that a first fault between a conductor and earth raises an alarm instead of tripping the supply in the middle of an operation. We plan the group 2 locations with the hospital, give each theatre or each group of beds a system of its own, place the panels, the alarm indicators and the socket outlets at each patient place, and design the equipotential bonding of each room, to IEC 60364-7-710 and HTM 06-01, or to NFPA 99 and NFPA 70 where a project follows American codes.
We supply the medical IT panels, the isolating transformers, the insulation monitoring devices and the alarm indicators from the brands we carry, and our teams install them in the theatres and the critical care units and connect them to the safety supplies. Each system is tested before the hospital takes the rooms into use: the insulation monitor's alarm at its set value, the transformer's leakage current and its load and temperature monitoring, and the resistance of every bonding connection. Under a service contract we then maintain the systems we installed, with the periodic tests IEC 60364-7-710 asks for.